• DocumentCode
    1456863
  • Title

    AC measurements of the minimum longitudinal resistance of a QHE sample from 10 Hz to 10 kHz

  • Author

    Piquemal, François P M ; Trapon, Gérard R. ; Genevès, Gérard P J

  • Author_Institution
    Lab. Central des Ind. Electriques, Fontenay aux Roses, France
  • Volume
    45
  • Issue
    6
  • fYear
    1996
  • fDate
    12/1/1996 12:00:00 AM
  • Firstpage
    918
  • Lastpage
    922
  • Abstract
    AC measurements of the longitudinal resistance, Rxx, of a quantum Hall effect (QHE) sample have been made in a frequency range from 10 Hz to 10 kHz. The results show no frequency effect on the minimum value of Rxx corresponding to the quantum numbers i=2 and i=4, within the measurement resolution of 0.5 mΩ. Therefore, the influence of frequency on the value of the quantized Hall resistance, RH, should not exceed a few parts in 109 . Some unwanted effects detected during the development of the resistance bridge have been pointed out
  • Keywords
    III-V semiconductors; electric resistance measurement; gallium arsenide; measurement standards; quantum Hall effect; 10 Hz to 10 kHz; AC measurement; GaAs; QHE sample; frequency effect; longitudinal resistance; quantized Hall resistance; quantum Hall effect; resistance bridge; Bridge circuits; Electric breakdown; Electrical resistance measurement; Frequency measurement; Gallium arsenide; Hall effect; Probes; Quantization; Quantum capacitance; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.543986
  • Filename
    543986