DocumentCode
1457514
Title
In vivo measurements of ultrasonic backscattering in blood
Author
Wang, Shyh-Hau ; Shung, K. Kirk
Author_Institution
Dept. of Biomed. Eng., Chung Yuan Christian Univ., Chung Li, Taiwan
Volume
48
Issue
2
fYear
2001
fDate
3/1/2001 12:00:00 AM
Firstpage
425
Lastpage
431
Abstract
Ultrasonic backscattering in blood including its dependence on the hematocrit, plasma proteins, shear rate, and flow disturbance, has been studied extensively theoretically and experimentally in vitro. However, much of the result has never been validated in vivo. To do so, backscattering measurements were made on pigs using a 10-MHz non-focused intravascular transducer in direct contact with blood. The probe was placed in either the abdominal aorta or the inferior vena cava. The backscattering coefficient (BSC) of blood flowing in these vessels as well as downstream from a stenosis was measured using an approach that was originally developed for measurements with focused transducers. With this approach, 6% porcine red cell saline suspensions prepared immediately after each in vivo measurement were used as the reference medium. Result from seven pigs at hematocrits ranging from 29 to 36% (31.9±2.5%) demonstrated that BSC of blood in the vena cava, (4.62±2.06)×10 -5 cm-sr -1, is consistently higher than that in the aorta, (2.65±1.22)×10 -5 cm-sr -1. The difference has been attributed to the lower shear rate and the formation of red cell aggregation in venous blood. These in vivo results are in agreement with those obtained in vitro. In response to stenoses created by ligating the aorta, backscattering of the blood measured downstream from the stenosis showed that the closer the site of measurement relative to the stenosis, the higher the backscatter, presumably resulting from the higher degree of flow disturbance. In vitro backscattering results on porcine whole blood were also acquired at 20 MHz with a Diasonics intravascular scanner.
Keywords
backscatter; bioacoustics; biomedical ultrasonics; blood; blood flow measurement; ultrasonic scattering; ultrasonic velocity measurement; 10 MHz; abdominal aorta; backscattering coefficient; blood; flow disturbance; hematocrit; in vivo measurements; inferior vena cava; nonfocused intravascular transducer; pigs; plasma proteins; porcine red cell saline suspensions; red cell aggregation; shear rate; stenosis; ultrasonic backscattering; venous blood; Backscatter; Blood; Fluid flow measurement; In vitro; In vivo; Plasma measurements; Probes; Proteins; Transducers; Ultrasonic variables measurement; Animals; Aorta, Abdominal; Aortic Valve Stenosis; Biomedical Engineering; Blood; Hematocrit; Scattering, Radiation; Swine; Venae Cavae;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.911725
Filename
911725
Link To Document