Abstract
A small prototype of gamma-ray detector suitable for both imaging and spectroscopy, based on a single CsI(Tl) scintillator coupled to a two-dimensional Si drift detector (SDD) array, has been developed. The determination of the position of interaction is performed by applying the centroid method to the signals from the single SDD units, while the measurement of the γ photon energy is performed by summing together these signals. This first prototype has been developed in order to evaluate the performances of a CsI(Tl)-SDD detection scheme in view of the realization of larger scintillation cameras for nuclear medicine. The low electronics noise of the SDD allows obtaining good performances both in position and in energy resolution. A position resolution of the order of 0.6 mm and an energy resolution of about 17% FWHM at 122 keV have been measured
Keywords
gamma-ray detection; position sensitive particle detectors; silicon radiation detectors; solid scintillation detectors; CsI(Tl) scintillator; CsI:Tl; Si; Si drift detector array; centroid method; energy resolution; gamma-camera; gamma-ray detector; position resolution; Energy measurement; Energy resolution; Gamma ray detection; Gamma ray detectors; Optical imaging; Performance evaluation; Position measurement; Prototypes; Sensor arrays; Spectroscopy;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.914472
Filename
914472
Link To Document