• DocumentCode
    1464447
  • Title

    Influence of Different Configurations of Nonideal Calibration Standards on Vector Network Analyzer Performance

  • Author

    Stumper, Ulrich ; Schrader, Thorsten

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Braunschweig, Germany
  • Volume
    61
  • Issue
    7
  • fYear
    2012
  • fDate
    7/1/2012 12:00:00 AM
  • Firstpage
    2034
  • Lastpage
    2041
  • Abstract
    To perform system error correction of a microwave vector network analyzer (VNA), several calibration procedures are applied, several of which are derived from the well-known short-open-load-through (SOLT) calibration method where lumped one-port circuits like short circuits, open circuits, and low-reflective loads are used as calibration standards. These derived calibration algorithms only use three calibration standard connections-instead of six for SOLT; hence, the calibration time is considerably reduced. Various configurations of high- and low-reflective standards at test ports 1 and 2 of the VNA are possible. Using inaccurate calibration standard modeling, the VNA calibration is incorrect, and hence, inaccurate -parameters of a device under test (DUT) are measured. In this paper, it is theoretically and experimentally shown how, due to non-ideal calibration standard modeling, the deviations of measured -parameters of the DUT from their hardware-defined true values depend on the inaccurate calibration standard modeling, on the configurations of the calibration standards at the test ports, on the reflection values of the calibration standards, and on the -parameter values of the DUTs themselves.
  • Keywords
    S-parameters; calibration; microwave circuits; microwave measurement; network analysers; DUT measurement; S-parameter; SOLT calibration method; device under test measurement; hardware-defined true value; high-reflective standard; low-reflective load; low-reflective standard; lumped one-port circuit; microwave VNA; microwave vector network analyzer; nonideal calibration standard modelling; open circuit; short circuit; short-open-load-through calibration method; system error correction; Attenuators; Calibration; Equations; Mathematical model; Scattering parameters; Sensitivity; Standards; $S$-parameters; Calibration standards; VNA calibration; microwave measurements; sensitivity coefficients; uncertainty of measurement; vector network analyzer (VNA);
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2012.2182889
  • Filename
    6165355