Title :
Bidirectional pixel based focusing in conventional B-mode ultrasound imaging
Author :
Bae, Moo-Ho ; Jeong, Mok-Kun
Author_Institution :
Medison Co. Ltd., South Korea
Volume :
34
Issue :
22
fYear :
1998
fDate :
10/29/1998 12:00:00 AM
Firstpage :
2105
Lastpage :
2107
Abstract :
The authors propose an all point transmit and receive focusing algorithm which combines transmit synthetic focusing and receive dynamic focusing in conventional B-mode imaging systems. Transmit focusing is synthesised using the RF data of each scanline, together with all other relevant scanline data previously stored. Computer simulation shows that the proposed method has better lateral resolution and higher SNR than conventional focusing methods at all imaging depth points
Keywords :
image scanners; ultrasonic focusing; ultrasonic imaging; ultrasonic transducer arrays; B-mode ultrasound imaging; SNR; all point transmit and receive focusing algorithm; bidirectional pixel based focusing; imaging depth points; lateral resolution; receive dynamic focusing; scanline RF data; transmit synthetic focusing;