DocumentCode :
1467721
Title :
Adaptive strain estimation using retrospective processing [medical US elasticity imaging]
Author :
Lubinski, Mark A. ; Emelianov, Stanislav Y. ; O´Donnell, Matthew
Author_Institution :
Dept. of Biomed. Eng. & Electr. Eng., Michigan Univ., Ann Arbor, MI, USA
Volume :
46
Issue :
1
fYear :
1999
Firstpage :
97
Lastpage :
107
Abstract :
Because errors in displacement and strain estimates depend on the magnitude of the induced strain, the strain signal-to-noise ratio (SNR) will be a function of the applied deformation. If deformation is applied at the body surface, it is difficult during data acquisition to select a single surface displacement providing the highest strain SNR throughout the image. By applying continuous deformation and capturing data in real-time, the surface displacement providing the highest strain SNR can be selected retrospectively. A method to adaptively optimize strain SNR over the image plane using retrospective processing is presented and demonstrated with experimental results.
Keywords :
acoustic signal processing; adaptive signal processing; biomedical ultrasonics; data acquisition; adaptive strain estimation; applied deformation; continuous deformation; data acquisition; medical US elasticity imaging; retrospective processing; strain signal-to-noise ratio; surface displacement; Biomedical measurements; Capacitive sensors; Data acquisition; Displacement measurement; Elasticity; Optimization methods; Signal to noise ratio; Strain measurement; Surface reconstruction; Ultrasonic imaging;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/58.741428
Filename :
741428
Link To Document :