• DocumentCode
    1470809
  • Title

    Low-jitter trigger system for pulse calibration and intercomparison of high-speed samplers

  • Author

    Paulter, N.G.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • Volume
    47
  • Issue
    3
  • fYear
    1998
  • fDate
    6/1/1998 12:00:00 AM
  • Firstpage
    606
  • Lastpage
    608
  • Abstract
    A low-jitter (<1 ps) trigger system for pulse-waveform-based calibration and intercomparison of high-speed samplers is described. The system uses a commercially available pulse generator and pulse splitter/delay line
  • Keywords
    calibration; delay lines; jitter; oscilloscopes; pulse generators; recorders; signal sampling; trigger circuits; high-speed samplers; intercomparison; low-jitter trigger system; pulse calibration; pulse generator; pulse splitter/delay line; pulse-waveform-based calibration; Bandwidth; Calibration; Circuits; Delay lines; Jitter; Oscilloscopes; Pulse generation; Pulse measurements; Sampling methods; Space vector pulse width modulation;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.744308
  • Filename
    744308