Abstract
This paper discusses systematic errors in the AC-DC difference determination resulting from the comparison of thermal voltage converters with different input connectors. Tests of two asymmetrical tees, with GR-874 and type-N arms, as well as the internal tee of the NRC calorimetric thermal voltage converter, are presented
Keywords
calibration; convertors; measurement errors; transfer standards; voltage measurement; AC-DC difference determination; HF thermal voltage converters; asymmetrical tees; calibration; calorimetric thermal voltage converter; coaxial transmission lines; different input connectors; internal tee; standing wave error; systematic errors; Arm; Calibration; Coaxial components; Connectors; Electric variables measurement; Frequency conversion; Hafnium; Impedance; Voltage; Wavelength measurement;