NRC-NIST comparison of power meter calibrations at 60 Hz and ranges up to 600 V, 100 A
Author
So, Eddy ; Angelo, D. ; Nelson, Tom ; Snider, Lisa
Author_Institution
Nat. Res. Council of Canada, Ottawa, Ont., Canada
Volume
50
Issue
2
fYear
2001
fDate
4/1/2001 12:00:00 AM
Firstpage
353
Lastpage
355
Abstract
A number of international comparisons of power meter calibrations at power frequencies were conducted in the past, all of them being done at 120 V and 5 A. This comparison of power meter calibrations between the National Research Council of Canada and the National Institute of Standards and Technology, USA, was done to include voltage ranges up to 600 V, current ranges up to 100 A, and at power factors of 1.0, 0.5 lead and lag, and 0.0 lead and lag. The comparison was implemented by a transfer standard consisting of a modified commercial time-division multiplier type wattmeter. The results indicate that there are no significant differences in the overall accuracy of AC power measurements in each laboratory up to 600 V, 100 A, at all power factors
Keywords
calibration; measurement uncertainty; power measurement; power meters; transfer standards; wattmeters; 100 A; 60 Hz; 600 V; AC power measurements; NRC-NIST comparison; international comparisons; power frequencies; power meter calibrations; time-division multiplier type wattmeter; transfer standard; uncertainty; Calibration; Circuit testing; Councils; Frequency; Laboratories; NIST; Reactive power; Standards development; Voltage; Wattmeters;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.918140
Filename
918140
Link To Document