DocumentCode :
1471026
Title :
International comparison of a pulse calibrator used in high-voltage impulse calibration
Author :
Li, Yi ; Rungis, Juris ; McComb, Terry R. ; Lucas, Wolfgang
Author_Institution :
Nat. Meas. Lab., CSIRO, Lindfield, NSW, Australia
Volume :
50
Issue :
2
fYear :
2001
fDate :
4/1/2001 12:00:00 AM
Firstpage :
430
Lastpage :
435
Abstract :
Measurements on a pulse generator have been made at the NRC in Canada and the PTB in Germany. Most of the measured peak values (Vp ) among the laboratories agree to ±0.5%, the front times (T 1) to ±2.0%, and the times to half values (T2 ) to ±1.0%. The differences between calculated calibrator outputs and measured averages were found to be 0.13%, 0.63%, and 0.32% for Vp, T1, and T2, respectively. Contributions of impulse analysis software to the measurement errors are also discussed
Keywords :
calibration; high-voltage techniques; impedance matching; measurement errors; measurement standards; measurement uncertainty; pulse generators; pulse measurement; voltage measurement; calculated calibrator outputs; high-voltage impulse calibration; impulse analysis software; impulse measurement; international comparison; low source impedance; matched calibrator; measured averages; measurement errors; measurement uncertainty; peak voltage measurement; pulse calibrator; pulse generator; Australia; Calibration; Circuit simulation; Councils; Laboratories; Pulse generation; Pulse measurements; Resistors; Switches; Voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.918159
Filename :
918159
Link To Document :
بازگشت