• DocumentCode
    1471152
  • Title

    Accuracy evaluation of a cesium fountain primary frequency standard at NIST

  • Author

    Meekhof, D.M. ; Jefferts, S.R. ; Stepanovíc, M. ; Parker, T.E.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Boulder, CO, USA
  • Volume
    50
  • Issue
    2
  • fYear
    2001
  • fDate
    4/1/2001 12:00:00 AM
  • Firstpage
    507
  • Lastpage
    509
  • Abstract
    A new laser-cooled Cs fountain primary frequency standard has been constructed at the National Institute of Standards and Technology (NIST) in Boulder, CO. The new standard, NIST-F1, is described here, including details of systematic frequency biases for which frequency corrections are made. We have submitted two evaluations to the Bureau International des Poids et Mesures (BIPM), which have Type A (statistical) uncertainties less than 1.5×10-15 and Type B (systematic) uncertainties less than 0.9×10-15
  • Keywords
    caesium; frequency standards; measurement uncertainty; probability; spectral line shift; statistical analysis; BIPM; Bureau International des Poids et Mesures; Cs; Cs fountain; NIST; National Institute of Standards and Technology; frequency corrections; measurement uncertainties; primary frequency standard; statistical uncertainties; systematic uncertainties; Atom lasers; Atomic beams; Atomic measurements; Frequency conversion; Laser beams; Laser transitions; Masers; NIST; Saturation magnetization; Uncertainty;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.918178
  • Filename
    918178