    Optimization of a phased-array transducer for multiple harmonic imaging in medical applications: frequency and topology

  • Author

    Matte, Guillaume M. ; Van Neer, Paul L M J ; Danilouchkine, Mike G. ; Huijssen, Jacob ; Verweij, Martin D. ; de Jong, Nico

  • Author_Institution
    Biomed. Eng., Erasmus Med. Center, Rotterdam, Netherlands
  • Volume
    58
  • Issue
    3
  • fYear
    2011
  • fDate
    3/1/2011 12:00:00 AM
  • Firstpage
    533
  • Lastpage
    546
  • Abstract
    Second-harmonic imaging is currently one of the standards in commercial echographic systems for diagnosis, because of its high spatial resolution and low sensitivity to clutter and near-field artifacts. The use of nonlinear phenomena mirrors is a great set of solutions to improve echographic image resolution. To further enhance the resolution and image quality, the combination of the 3rd to 5th harmonics - dubbed the superharmonics - could be used. However, this requires a bandwidth exceeding that of conventional transducers. A promising solution features a phased-array design with interleaved low- and high-frequency elements for transmission and reception, respectively. Because the amplitude of the backscattered higher harmonics at the transducer surface is relatively low, it is highly desirable to increase the sensitivity in reception. Therefore, we investigated the optimization of the number of elements in the receiving aperture as well as their arrangement (topology). A variety of configurations was considered, including one transmit element for each receive element (1/2) up to one transmit for 7 receive elements (1/8). The topologies are assessed based on the ratio of the harmonic peak pressures in the main and grating lobes. Further, the higher harmonic level is maximized by optimization of the center frequency of the transmitted pulse. The achievable SNR for a specific application is a compromise between the frequency-dependent attenuation and nonlinearity at a required penetration depth. To calculate the SNR of the complete imaging chain, we use an approach analogous to the sonar equation used in underwater acoustics. The generated harmonic pressure fields caused by nonlinear wave propagation were modeled with the iterative nonlinear contrast source (INCS) method, the KZK, or the Burger´s equation. The optimal topology for superharmonic imaging was an interleaved design with 1 transmit element per 6 receive elements. It improves the SNR by ~5 dB compared w- th the interleaved (1/2) design reported in literature. The optimal transmit frequency for superharmonic echocardiography was found to be 1.0 to 1.2 MHz. For superharmonic abdominal imaging this frequency was found to be 1.7 to 1.9 MHz. For 2nd-harmonic echocardiography, the optimal transmit frequency of 1.8 MHz reported in the literature was corroborated with our simulation results.
  • Keywords
    biomedical transducers; biomedical ultrasonics; harmonic generation; iterative methods; nonlinear acoustics; ultrasonic imaging; ultrasonic transducer arrays; Burger´s equation; INCS method; KZK equation; Khokhlov-Zabolotskaya-Kuznetsov equation; diagnostic echographic systems; echographic image resolution; frequency 1.0 MHz to 1.2 MHz; frequency 1.7 MHz to 1.9 MHz; frequency dependent attenuation; harmonic peak pressure ratio; harmonic pressure fields; high frequency elements; image quality enhancement; imaging resolution enhancement; iterative nonlinear contrast source method; low frequency elements; medical applications; multiple harmonic imaging; nonlinear acoustic wave propagation; nonlinear phenomena mirrors; optimal transmit frequency; penetration depth; phased array design; phased array transducer optimisation; receiving aperture element number optimisation; receiving aperture topology optimisation; second harmonic echocardiography; second harmonic imaging; sonar equation analogue; superharmonic abdominal imaging; superharmonic echocardiography; superharmonic imaging topology; transmitted pulse center frequency optimisation; Arrays; Equations; Harmonic analysis; Imaging; Mathematical model; Topology; Transducers; Acoustics; Computer Simulation; Echocardiography; Humans; Signal Processing, Computer-Assisted; Transducers;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2011.1837
  • Filename
    5733256