Abstract
We describe a cryogenic radiometric calibration system suitable for high sensitivity, low background power bolometers. The system uses a variable temperature blackbody source capable of temperatures of 4 to 20 K, whose output covers the sub-millimeter wavelength regime. The detector is an eight-channel, filled-focal-plane array of superconducting transition edge sensor (TES) bolometers. The system is designed to test components of the optical system such as polarizers, filters and stray light suppressors as well as the performance of the detector array elements. The bolometer response to incident submillimeter radiation is measured and compared to the calculated output of the source seen by the bolometer
Keywords
bolometers; radiometry; superconducting devices; detector array elements; filled-focal-plane array; filters; high sensitivity low background power bolometers; incident submillimeter radiation; polarizers; stray light suppressors; sub-millimeter wavelength regime; sub-picowatt precision radiometry; superconducting transition edge sensor bolometers; variable temperature blackbody source; Bolometers; Calibration; Cryogenics; Detectors; Optical arrays; Optical filters; Radiometry; Sensor arrays; Superconducting filters; Temperature sensors;
fLanguage
English
Journal_Title
Applied Superconductivity, IEEE Transactions on
Publisher
ieee
ISSN
1051-8223
Type
jour
DOI
10.1109/77.919415
Filename
919415
Link To Document