Title :
Antenna performance for localization of capsule endoscope
Author :
Ara, Perzila ; Heimlich, Michael ; Dutkiewicz, Eryk
Abstract :
Recent developments in capsule endoscopy have highlighted the need for accurate techniques to estimate the location of a capsule endoscope in the gastrointestinal (GI) tract. Besides the channel propagation environment that can highly influence the accuracy of the capsule location estimation, the deployed antenna properties and its radiation pattern can also have a significant impact on the performance of the location estimation technique. Therefore, it is essential to select the antenna type that can meet the location estimation requirements. In this paper, the studies are done for two different antenna types: half-dipole and loop at the frequency of 2.4 GHz to explore and compare the effect of different antenna types. The results show that the half-dipole antenna outperforms the loop antenna in respect to the signal attenuation inside the human abdomen region.
Keywords :
biological organs; dipole antennas; endoscopes; loop antennas; antenna performance; antenna properties; capsule endoscope localization estimation; capsule endoscopy; channel propagation environment; frequency 2.4 GHz; gastrointestinal tract; half-dipole antenna; human abdomen region; location estimation requirements; loop antenna; radiation pattern; signal attenuation; Antenna radiation patterns; Dipole antennas; Intestines; Radar; Radar antennas; Capsule endoscope; Radar; half-dipole antenna; loop antenna;
Conference_Titel :
Medical Information and Communication Technology (ISMICT), 2014 8th International Symposium on
Conference_Location :
Firenze
DOI :