DocumentCode :
1477328
Title :
Observation and cancellation of a perturbing dc stark shift in strontium optical lattice clocks
Author :
Lodewyck, Jérôme ; Zawada, Michal ; Lorini, Luca ; Gurov, Mikhail ; Lemonde, Pierre
Author_Institution :
Obs. de Paris, Univ. Pierre et Marie Curie (UPMC), Paris, France
Volume :
59
Issue :
3
fYear :
2012
fDate :
3/1/2012 12:00:00 AM
Firstpage :
411
Lastpage :
415
Abstract :
We report on the observation of a dc Stark frequency shift at the 10-13 level by comparing two strontium optical lattice clocks. This frequency shift arises from the presence of electric charges trapped on dielectric surfaces placed under vacuum close to the atomic sample. We show that these charges can be eliminated by shining UV light on the dielectric surfaces, and characterize the residual dc Stark frequency shift on the clock transition at the 10-18 level by applying an external electric field. This study shows that the dc Stark shift can play an important role in the accuracy budget of lattice clocks, and should be duly taken into account.
Keywords :
atomic clocks; electric charge; optical lattices; strontium; Sr; dielectric surface; electric charge; electric field; optical lattice clock; perturbing DC Stark frequency shift cancellation; perturbing DC Stark frequency shift observation; residual DC Stark frequency shift; shining UV light elimination; Atom optics; Atomic clocks; Cavity resonators; Electric fields; Lattices; Mirrors;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2012.2209
Filename :
6174185
Link To Document :
بازگشت