DocumentCode :
1479005
Title :
Simulation of phase effects in magnetic resonance flow imaging
Author :
Singh, Manbir ; Oshio, Koichi
Author_Institution :
Dept. of Radiol., Univ. of Southern California, Los Angeles, CA, USA
Volume :
37
Issue :
3
fYear :
1990
fDate :
6/1/1990 12:00:00 AM
Firstpage :
1316
Lastpage :
1320
Abstract :
Using a phantom containing stationary and flowing regions, the phase-related effects of flow on spin echo and echo planar images have been investigated. The role of parameters such as flow direction, mean velocity, velocity modulation, gradient pulses, echo time and data sampling time in producing artifacts and the reduction of these artifacts by a flow compensating pulse have been examined. It is shown that, due to the finite data sampling window, the flow compensating pulse cannot correct fully for phase shifts even for first-order motion
Keywords :
biomedical NMR; flow visualisation; haemodynamics; spin echo (NMR); data sampling time; echo planar images; echo time; flow direction; gradient pulses; magnetic resonance flow imaging; mean velocity; phase effects; spin echo; velocity modulation; Image sampling; Imaging phantoms; Magnetic resonance; Magnetic resonance imaging; Magnetization; Nuclear magnetic resonance; Protons; Pulse modulation; Radiology; Visualization;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.57382
Filename :
57382
Link To Document :
بازگشت