• DocumentCode
    1491082
  • Title

    Localization of the site of origin of cardiac activation by means of a heart-model-based electrocardiographic imaging approach

  • Author

    Li, Guanglin ; He, Benteng

  • Author_Institution
    Dept. of Electr. Eng. & Comput. Sci., Illinois Univ., Chicago, IL, USA
  • Volume
    48
  • Issue
    6
  • fYear
    2001
  • fDate
    6/1/2001 12:00:00 AM
  • Firstpage
    660
  • Lastpage
    669
  • Abstract
    The authors have developed a new approach to solve the inverse problem of electrocardiography in terms of heart model parameters. The inverse solution of the electrocardiogram (ECG) inverse problem is defined, in the present study, as the parameters of the heart model, which are closely related to the physiological and pathophysiological status of the heart, and is estimated by using an optimization system of heart model parameters, instead of solving the matrix equation relating the body surface ECGs and equivalent cardiac sources. An artificial neural network based preliminary diagnosis system has been developed to limit the searching space of the optimization algorithm and to initialize the model parameters in the computer heart model. The optimal heart model parameters were obtained by minimizing the objective functions, as functions of the observed and model-generated body surface ECGs. The authors have tested the feasibility of the newly developed technique in localizing the site of origin of cardiac activation using a pace mapping protocol. The present computer simulation results show that, the present approach for localization of the site of origin of ventricular activation achieved an averaged localization error of about 3 mm [for 5-μV Gaussian white noise (GWN)] and 4 mm (for 10-μV GWN), with standard deviation of the localization errors of being about 1.5 mm. The present simulation study suggests that this newly developed approach provides a robust inverse solution, circumventing the difficulties of the ECG inverse problem, and may become an important alternative to other ECG inverse solutions.
  • Keywords
    biomedical imaging; electrocardiography; inverse problems; neural nets; optimisation; physiological models; white noise; 10 muV; 5 muV; artificial neural network based preliminary diagnosis system; body surface ECGs; cardiac activation site origin localization; electrodiagnostics; equivalent cardiac sources; heart model parameters; heart-model-based electrocardiographic imaging approach; localization errors; matrix equation relating; optimization system; pace mapping protocol; pathophysiological status; physiological status; Artificial neural networks; Computer errors; Computer networks; Computer simulation; Electrocardiography; Equations; Heart; Inverse problems; Protocols; Testing; Algorithms; Body Surface Potential Mapping; Cardiac Pacing, Artificial; Computer Simulation; Electrocardiography; Heart Conduction System; Heart Ventricles; Humans; Mathematics; Models, Cardiovascular; Neural Networks (Computer);
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.923784
  • Filename
    923784