• DocumentCode
    1492722
  • Title

    Microwave TDR for Real-Time Control of Intravenous Drip Infusions

  • Author

    Cataldo, Andrea ; Cannazza, Giuseppe ; Giaquinto, Nicola ; Trotta, Amerigo ; Andria, Gregorio

  • Author_Institution
    Dept. of Eng. for Innovation, Univ. of Salento, Lecce, Italy
  • Volume
    61
  • Issue
    7
  • fYear
    2012
  • fDate
    7/1/2012 12:00:00 AM
  • Firstpage
    1866
  • Lastpage
    1873
  • Abstract
    This paper explores the use of a microwave-reflectometry-based system for the automatic control and real-time monitoring of the flow and of the liquid level in intravenous (IV) medical infusions. In medical and hospital contexts, other kinds of devices, mainly based on the optical detection and counting of the infusion drops, are used. Nevertheless, the proposed system is aimed at circumventing some typical drawbacks deriving from the adoption of these traditional methods, thus allowing an efficient alternative for automatically monitoring the instantaneous flow of IV medical solutions. To this purpose, the proposed system combines microwave time-domain reflectometry (TDR) measurements with a noninvasive sensing element (i.e., strip electrodes directly attached to the external surface of the infusion bottle). Experimental results confirm that, by using low-cost portable TDR devices, the solution flow process can be controlled with acceptable accuracy. Therefore, the proposed method can be regarded as a promising control tool for in-hospital patient management as well as for telemedicine programs.
  • Keywords
    biomedical electrodes; biomedical measurement; flow control; level control; medical control systems; microwave reflectometry; patient monitoring; sensors; time-domain reflectometry; IV medical solutions; automatic flow control; hospital contexts; in-hospital patient management; infusion bottle; infusion drops; intravenous drip infusions; intravenous medical infusions; liquid level; low-cost portable TDR devices; medical contexts; microwave TDR; microwave time-domain reflectometry measurements; microwave-reflectometry-based system; noninvasive sensing element; optical detection; real-time control; real-time monitoring; solution flow process; strip electrodes; telemedicine programs; Accuracy; Microwave measurements; Monitoring; Probes; Sensors; Volume measurement; Biomedical monitoring; medical services; microwave measurements; microwave reflectometry; time-domain reflectometry (TDR);
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2012.2192346
  • Filename
    6182713