DocumentCode :
1494135
Title :
An FPGA-Integrated Time-to-Digital Converter Based on Two-Stage Pulse Shrinking
Author :
Szplet, Ryszard ; Klepacki, Kamil
Author_Institution :
Mil. Univ. of Technol., Warsaw, Poland
Volume :
59
Issue :
6
fYear :
2010
fDate :
6/1/2010 12:00:00 AM
Firstpage :
1663
Lastpage :
1670
Abstract :
We present the design and test results of a new time-to-digital converter based on the cyclic pulse shrinking method and implemented in a field-programmable gate array (FPGA) device. The pulse shrinking is realized in a loop containing two complementary delay lines. The first line shrinks, and the second line stretches the duration time of a circulating pulse; hence, the length ratio of the lines determines the pulse-shrinking capability of the converter. This resolution control mechanism is different from the bias adjustment typically used in complementary metal-oxide-semiconductor (CMOS) application-specific integrated circuit (ASIC) pulse-shrinking elements. Two forms of representation of a measured time interval (the pulse width and the time interval between two pulses) are utilized in the described converter to improve the efficiency of resolution control. To diminish the jitter of the edges of a circulating pulse and, consequently, to increase the precision of the converter, a two-stage conversion is introduced. The first stage, having a low resolution, rapidly shortens the measured pulse, thus limiting the number of cycles, whereas the second stage provides a final high resolution within a narrow time interval range. The optimal resolution of the first conversion stage is theoretically derived. Mostly, the same logical resources of the FPGA device are utilized in both conversion stages; thus, the overall area overhead of the converter is reduced. The converter has a resolution of 42 ps and a measurement uncertainty of below 56 ps within the measurement range of 11.5 ns. The converter has been implemented in a general-purpose reprogrammable device Spartan-3 (Xilinx).
Keywords :
analogue-digital conversion; circuit testing; delay lines; field programmable gate arrays; jitter; FPGA-integrated time-to-digital converter; application-specific integrated circuit; complementary delay lines; complementary metal-oxide-semiconductor; field-programmable gate array device; general-purpose reprogrammable device Spartan-3; jitter; resolution control mechanism; time 11.5 ns; time 42 ps; two-stage conversion; two-stage pulse shrinking method; Field-programmable gate arrays; pulse shrinking; time digitizer; time-to-digital converter; two-stage interpolation;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2009.2027777
Filename :
5280372
Link To Document :
بازگشت