Dynamic frame pairing in real-time freehand elastography
Author :
Rongmin Xia ; Guozhi Tao ; Thittai, Arun
Author_Institution :
Dept. of Radiol., Univ. of Texas at Houston, Houston, TX, USA
Volume :
61
Issue :
6
fYear :
2014
fDate :
Jun-14
Firstpage :
979
Lastpage :
985
Abstract :
Quasi-static ultrasound (US) elastography is now a well-established technique that involves acquiring US (RF/envelope) signals from an imaging plane before and after a small quasi-static compression to form axial strain elastograms (ASE). The image quality of the ASEs is a function of the applied axial strain. This relationship was extensively investigated and formalized in terms of strain filter in the literature. Most of the work in elastography formed elastograms by choosing pre- and post-compression frames separated by a desired compression strain. Although this approach is feasible in simulations and in vitro/in vivo experiments that involve controlled compression, it has been a challenge to do this during freehand compression in real time. In this work, we describe a one-prediction-one- correction method that dynamically selects pre- and post- compression frames to form an elastogram, based on the applied axial strain level. We validate the method using controlled compression experiments on phantoms and compare the performance of the dynamic frame pairing method against successive-frame pairing method in terms of the contrast-to-noise ratio (CNRe). Further, we demonstrate the advantages of the new method with the help of freehand acquired data from phantom experiments and in vivo breast data. The results demonstrate that the frame-pairing identified by the dynamic method matched the frame pairing that was designed to yield an applied axial strain of ~1%. The CNRe obtained by the traditional approach varied from as low as ~5 to as high as ~25, depending on the choice of skip number and compression rate. However, the dynamic frame pairing method provided elastograms with a CNRe that was consistently around ~20, irrespective of the compression rate. The results from analysis of 22 in vivo breast data demonstrated that the dynamic pairing method generated elastograms such that the frame-average axial strain (FAAS) of each frame in- the cineloop is consistently ~1% (0.011 ± 0.001).
Keywords :
biomedical ultrasonics; data compression; elasticity; image coding; image matching; medical image processing; noise; phantoms; contrast-to-noise ratio; dynamic frame pairing method; frame-average axial strain elastograms; image matching; image quality; in vivo breast data; one-prediction-one- correction method; phantom experiments; quasistatic compression strain; quasistatic ultrasound elastography; strain filter; FAA; Image coding; In vivo; Phantoms; Radio frequency; Real-time systems; Strain;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2014.2993
Filename :
6819213
Link To Document :
بازگشت