Development of capacitance measurement system for human serum albumin detection
Author
Tsai, J.-Z. ; Chen, Chuan-Jie ; Liu, J.-T. ; Hsin, Yue-Ming ; Chen, Wei-Yu ; Hsueh, K.-H.
Author_Institution
Dept. of Electr. Eng., Nat. Central Univ., Jhongli, Taiwan
Volume
46
Issue
10
fYear
2010
Firstpage
678
Lastpage
680
Abstract
The use of interdigitated electrodes and binding with Cibacron blue F3GA by self-assembly technology, which has high affinity for human serum albumin (HSA), and serves as a receptor for HSA detection, is reported. In preliminary tests, the interdigitated chip measured the impedance of HSA by electrochemical impedance spectroscopy and this result indicated the characteristic frequency of impedance change below 1.1 kHz. However, a portable capacitance measurement system made using a standard operational amplifier circuit and AC excitation was applied at 1 kHz. The detection result showed a significant linear correlation between HSA concentration (0.1 to 1 mg/ml) and output voltage. The regression equation was Y 17.91 323.72X with R2 = 0.999. This reported study has successfully demonstrated that the capacitive sensor has great potential for applications in HSA detection.
Keywords
biomedical electrodes; biomedical measurement; capacitive sensors; electrochemical electrodes; electrochemical impedance spectroscopy; enzymes; kidney; molecular biophysics; regression analysis; AC excitation; capacitance measurement system; capacitive sensor; cibacron blue; electrochemical impedance spectroscopy; human serum albumin detection; interdigitated electrodes; regression equation; self-assembly technology; standard operational amplifier circuit;
fLanguage
English
Journal_Title
Electronics Letters
Publisher
iet
ISSN
0013-5194
Type
jour
DOI
10.1049/el.2010.0234
Filename
5466338
Link To Document