DocumentCode :
1494464
Title :
A Large Field of View Coincidence Imaging System Based on One-Dimensional Sharing Block Detectors
Author :
Yamamoto, S. ; Matsumoto, K. ; Senda, M.
Author_Institution :
Kobe City Coll. of Technol., Kobe, Japan
Volume :
56
Issue :
5
fYear :
2009
Firstpage :
2672
Lastpage :
2677
Abstract :
For such as breast cancer imaging, dedicated coincidence imaging detector will be promising for new instrumentation of nuclear medicine. For this purpose, a large field of view (FOV) coincidence imaging system was developed and tested. The imaging system consists of four sets of one-dimensional sharing block detectors and a single gamma probe and measures coincidence between these two detectors. The one-dimensional sharing detector uses 5 GSO blocks optically coupled to 6 dual photomultiplier tubes (PMT). The size of a single GSO is 2.9 mm times 2.9 mm times 15 mm. The signals from four one-dimensional sharing detectors are fed to weighted summing circuits. They are digitized by 100 MHz free running analog to digital (A-D) converters and are used to calculate the position of gamma interaction. Coincidence was measured between this gamma imaging detector (panel detector) and a hand-held single gamma probe. Acquired data was transferred to a personal computer and displayed. Most spots corresponding to GSO crystals were resolved in the position histogram. Spatial resolution was around 4 to 6 mm FWHM when the detector is greater than 25 mm. The sensitivity was 0.1-0.2% when the single probe was 25 mm away from the imager surface. Images of phantoms were successfully obtained. We conclude developed coincidence imaging system has large FOV and easy to handle, reasonable spatial resolution and may have potential to be a new instrument for nuclear medicine.
Keywords :
analogue-digital conversion; biomedical imaging; biomedical measurement; cancer; coincidence techniques; gamma-ray detection; phantoms; photomultipliers; radioisotope imaging; solid scintillation detectors; 6 dual photomultiplier tubes; FOV; GSO crystals; PMT; analog to digital converters; breast cancer imaging; coincidence imaging system; field of view; frequency 100 MHz; gamma imaging detector; gamma probe; nuclear medicine instrumentation; one-dimensional sharing block detectors; phantoms; spatial resolution; Breast cancer; Cancer detection; Gamma ray detection; Gamma ray detectors; Instruments; Nuclear medicine; Optical imaging; Probes; Spatial resolution; System testing; Block detector; GSO; concidence; position seisitive detector;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2009.2024120
Filename :
5280492
Link To Document :