DocumentCode :
1495246
Title :
A new method for gain correction in ultrasonic imaging
Author :
Richard, B.
Author_Institution :
Lab. de Biophys. Fac., Cochin, Paris, France
Volume :
10
Issue :
1
fYear :
1991
fDate :
3/1/1991 12:00:00 AM
Firstpage :
103
Abstract :
A system is proposed in which the depth for each ultrasonic image pixel, which is needed for scan conversion, is simultaneously given with the pixel coming out from the digital scan processor. The gain memory is only a line memory with less circuitry and is updated faster when the gain function is modified by the user. The use of the system in routine clinical work shows that post-gain correction capability is an improvement in terms of ease-of-use and image quality optimization
Keywords :
acoustic imaging; biomedical ultrasonics; digital scan processor; gain correction; gain memory; image quality optimization; medical diagnostic imaging; pixel; routine clinical work; scan conversion; ultrasonic imaging; Acoustic imaging; Circuits; Image converters; Image quality; Image storage; Pixel; Ultrasonic imaging;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.75617
Filename :
75617
Link To Document :
بازگشت