    Sequential Activation of Ground Pads Reduces Skin Heating During Radiofrequency Tumor Ablation: In Vivo Porcine Results

  • Author

    Schutt, David J. ; Swindle, M. Michael ; Helke, Kristi L. ; Bastarrika, Gorka ; Schwarz, Florian ; Haemmerich, Dieter

  • Author_Institution
    Div. of Pediatric Cardiology, Med. Univ. of South Carolina, Charleston, SC, USA
  • Volume
    57
  • Issue
    3
  • fYear
    2010
  • fDate
    3/1/2010 12:00:00 AM
  • Firstpage
    746
  • Lastpage
    753
  • Abstract
    Skin burns below ground pads during monopolar RF ablation are increasingly prevalent, thereby hindering the development of higher power RF generators capable of creating larger tumor ablation zones in combination with multiple or new applicators. Our goal was to evaluate reduction in skin temperatures via additional ground pads in an in vivo porcine model. Three ground pads placed on the animal´s abdomen were activated either simultaneously or sequentially, where activation timing was adjusted to equilibrate skin temperature below each pad. Thirteen RF ablations (n = 4 simultaneous at 300 W, n = 5 sequential at 300 W, and n = 4 sequential at 375 W) were performed for 12 min via two internally cooled cluster electrodes placed in the gluteus maximus of domestic swine. Temperature rise at each pad and burn degree as determined via histology were compared. Ablation zone size was determined via T2-weighted MRI. Maximum temperature rise was significantly higher with simultaneous activation than with either of the sequential activation group (21.4??C versus 8.1??C or 9.6??C, p < 0.01). Ablation zone diameters during simultaneous (300 W) and sequential activations (300 and 375 W) were and 6.9 ?? 0.3, 5.6 ?? 0.3, and 7.5 ?? 0.6 cm, respectively. Sequential activation of multiple ground pads results in significantly lower skin temperatures and less severe burns, as measured by histological examination.
  • Keywords
    biomedical MRI; biomedical electrodes; biothermics; physiological models; radiation therapy; radiofrequency heating; skin; tumours; RF generators; T2-weighted MRI; abdomen; burn degree; domestic swine; gluteus maximus; ground pads; histological examination; in vivo porcine results; internally cooled cluster electrodes; monopolar RF ablation; power 300 W; power 375 W; radiofrequency tumor ablation; sequential activation; skin burns; skin heating; skin temperatures; temperature rise; time 12 min; tumor ablation zones; Abdomen; Applicators; Cogeneration; Heating; In vivo; Land surface temperature; Power generation; Radio frequency; Skin neoplasms; Timing; RF ablation; skin burns; tumor ablation; Algorithms; Animals; Burns; Catheter Ablation; Histocytochemistry; Magnetic Resonance Imaging; Skin; Skin Temperature; Swine;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2009.2033385
  • Filename
    5282635