DocumentCode :
1497870
Title :
Improved measurement accuracy in heterodyne laser interferometer using WDF
Author :
Lee, Wei-Jen ; Lee, Jeyull ; You, Keyou
Author_Institution :
Sch. of Inf. & Commun., Sungkyunkwan Univ., Suwon, South Korea
Volume :
45
Issue :
21
fYear :
2009
Firstpage :
1085
Lastpage :
1087
Abstract :
The heterodyne laser interferometer is restricted by some unwanted nonlinear and environmental errors which are caused mainly from the frequency-mixing and the non-uniformal air refraction, respectively. Such errors act as an obstacle for measurement accuracy, especially in nanometer scale. Proposed is a compensation algorithm based on the WDF (weighted least square: WLS and data fusion: DF) method to improve measurement accuracy. The WLS method has a capability of minimal reduced error to eliminate the nonlinearity error in a laser interferometer. In the DF method, two laser heads used to obtain multiple measurement data are combined together to reduce environmental errors.
Keywords :
heterodyne detection; light interferometers; light refraction; measurement errors; remote sensing by laser beam; sensor fusion; compensation algorithm; data fusion; environmental errors; frequency-mixing; heterodyne laser interferometer; laser heads; measurement accuracy; minimal reduced error; nonlinear errors; nonuniformal air refraction; weighted least square;
fLanguage :
English
Journal_Title :
Electronics Letters
Publisher :
iet
ISSN :
0013-5194
Type :
jour
DOI :
10.1049/el.2009.1427
Filename :
5284464
Link To Document :
بازگشت