• DocumentCode
    1498122
  • Title

    A 5 Telsa imaging magnet for imaging laboratory animals

  • Author

    Carolan, James L. ; Burns, W.A. ; Green, Michael A.

  • Author_Institution
    NCC Corp., Martinez, CA, USA
  • Volume
    25
  • Issue
    2
  • fYear
    1989
  • fDate
    3/1/1989 12:00:00 AM
  • Firstpage
    1759
  • Lastpage
    1762
  • Abstract
    A report is presented on the construction of the first of a series of magnetic resonance imaging (MRI) imaging magnets for laboratory animals. The magnet has a 33-cm warm bore with a design central induction of 5.5 T without active shielding and 5.0 T with active shielding. The magnet will be used for both imaging and spectroscopy of living animals. The active shield system is designed so that the 5-G line is less than 3 m from the magnet center when the magnet operates at design field. This permits the magnet to be used within an experimental space commonly available within a university building
  • Keywords
    biological techniques and instruments; biomagnetism; superconducting magnets; 33 cm; 5 T; 5-G line; MRI magnets; active shield system; active shielding; bore; construction; design central induction; external field reduction; imaging laboratory animals; imaging magnet; living animals imaging; spectroscopy of living animals; superconducting magnets; university building; Animals; Boring; Coils; Laboratories; Magnetic confinement; Magnetic moments; Magnetic resonance imaging; Magnetic shielding; Nuclear magnetic resonance; Spectroscopy;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.92641
  • Filename
    92641