• DocumentCode
    1508533
  • Title

    A finite element analysis of the load transfer between an above-knee residual limb and its prosthetic socket-roles of interface friction and distal-end boundary conditions

  • Author

    Zhang, Ming ; Mak, Arthur F T

  • Author_Institution
    Rehabilitation Eng. Center, Hong Kong Polytech. Univ., Kowloon, Hong Kong
  • Volume
    4
  • Issue
    4
  • fYear
    1996
  • fDate
    12/1/1996 12:00:00 AM
  • Firstpage
    337
  • Lastpage
    346
  • Abstract
    A nonlinear finite element model was developed to investigate the manner of the load transfer between the above-knee residual limb and its prosthetic socket. The model was two-dimensional (2-D) with consideration of nonlinear boundary conditions and large deformation. Interface elements were used to simulate the friction/slip conditions between skin and the socket. Different distal-end conditions, including air cushion distal-end loading, were simulated to explore their influences on the load transfer
  • Keywords
    artificial limbs; biomechanics; finite element analysis; friction; skin; above-knee residual limb; air cushion distal-end loading; distal-end boundary conditions; friction/slip conditions; interface elements; interface friction; large deformation; load transfer; nonlinear boundary conditions; nonlinear finite element model; prosthetic socket; skin; two-dimensional model; Biological tissues; Boundary conditions; Deformable models; Finite element methods; Friction; Knee; Prosthetics; Residual stresses; Skin; Sockets;
  • fLanguage
    English
  • Journal_Title
    Rehabilitation Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1063-6528
  • Type

    jour

  • DOI
    10.1109/86.547935
  • Filename
    547935