Title :
Measurement uncertainty assessment of the scanning laser acoustic microscope and application to canine skin and wound
Author :
Steiger, D.L. ; O´Brien, William D., Jr. ; Olerud, J.E. ; Riederer-henderson, Mary Ann ; Odland, George F.
Author_Institution :
Illinois Univ., Urbana, IL, USA
Abstract :
Accuracy and precision were experimentally evaluated for the scanning laser acoustic microscope (SLAM) by measurements on homogeneous liquids of known ultrasonic properties. Using aqueous solutions of bovine serum albumin, the attenuation coefficient accuracy and precision are +or-12% and +or-15%, respectively. Using Dow Corning 710, a silicone oil, the speed accuracy and precision are +or-2.9% and +or-0.4%, respectively. Precision was assessed using duplicate samples of canine skin and wound tissue. The estimated precisions in the measurement of the attenuation coefficient and speed were +or-16% and +or-1.7%, respectively.<>
Keywords :
acoustic microscopy; biological techniques and instruments; laser beam applications; photoacoustic effect; Dow Corning 710; SLAM; attenuation coefficient; bovine serum albumin; canine skin; scanning laser acoustic microscope; silicone oil; speed; ultrasonic; wound; Acoustic measurements; Bovine; Liquids; Measurement uncertainty; Microscopy; Optical attenuators; Petroleum; Simultaneous localization and mapping; Skin; Ultrasonic variables measurement;
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on