DocumentCode :
1511570
Title :
Quality Assurance for Josephson Array Voltage Standards
Author :
Zhou, Yinzhu ; Chua, Sze Wey ; Lee, Jinni ; Shan, Yueyan
Author_Institution :
Nat. Metrol. Centre, Agency for Sci., Technol. & Res. (A*STAR), Singapore, Singapore
Volume :
60
Issue :
7
fYear :
2011
fDate :
7/1/2011 12:00:00 AM
Firstpage :
2578
Lastpage :
2583
Abstract :
A series of quality assurance (QA) procedures for the Josephson array voltage standard (JAVS) has been implemented to improve the efficiency and quality of dc voltage measurements in the National Metrology Centre, Agency for Science, Technology, and Research (A*STAR), Singapore. Periodic direct comparisons between two JAVS systems are required in the QA procedure to ensure accurate representation of the volt for calibration of secondary standards. QA-oriented setup and procedure are developed for better control of microwave power and locking frequency when using the millimeter-wave synthesizer. The analysis-of-variance method and the Monte Carlo method are implemented in data reporting and uncertainty evaluation processes.
Keywords :
measurement standards; voltage measurement; Josephson array voltage standards; Monte Carlo method; analysis-of-variance method; dc voltage measurements; locking frequency; microwave power; millimeter-wave synthesizer; quality assurance; Analysis of variance; Arrays; Measurement uncertainty; Standards; Synthesizers; Uncertainty; Voltage measurement; Direct comparison; Josephson arrays; measurement standards; quality assurance (QA); uncertainty; voltage;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2011.2139250
Filename :
5764538
Link To Document :
بازگشت