DocumentCode :
1512210
Title :
Novel gradient coil set with an interstitial gap for interventional nuclear magnetic resonance applications
Author :
Petropoulos, Labros S. ; Morich, Michael A.
Author_Institution :
NMR Div., Picker Int. Inc., Highland Heights, OH, USA
Volume :
33
Issue :
5
fYear :
1997
fDate :
9/1/1997 12:00:00 AM
Firstpage :
4107
Lastpage :
4109
Abstract :
The introduction of more open magnet geometries in Magnetic Resonance Imaging (MRI) system calls for the design of new gradient coils with similar shape as the corresponding main magnet. In this paper, a novel technique for designing minimum inductance gradient coils with an interstitial gap which is suitable for a cylindrical open magnet structure with two separated gaps is presented. Design examples are given for the axial and the transverse gradient coils. A comparison between the gap size and the stored magnetic energy for the same gradient strength indicates a near-quadratic relationship between these two quantities
Keywords :
NMR imaging; coils; cylindrical open magnet; gradient coil; inductance; interstitial gap; interventional nuclear magnetic resonance; magnetic energy storage; magnetic resonance imaging; Coils; Current density; Geometry; Magnetic fields; Magnetic resonance imaging; Magnetic separation; Minimally invasive surgery; Nuclear magnetic resonance; Shape; Superconducting magnets;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/20.619678
Filename :
619678
Link To Document :
بازگشت