Title :
Power efficient output stage for high density implantable stimulators
Author :
Sooksood, K. ; Ortmanns, Maurits
Author_Institution :
Dept. of Electron. Eng., King Mongkut´s Inst. of Technol., Bangkok, Thailand
Abstract :
In implantable stimulators, a current-controlled stimulation is preferred over a voltage-controlled one due to its safety. However, the power efficiency is a major disadvantage. By reducing the headroom voltage needed in the current driver, power efficiency of a constant current stimulation is improved. A promising technique is to bias the transistor in the triode region whereby improving output impedance through the regulated cascode structure. This comes with a feature of an implicit compliance monitor. Proposed is a new power efficient high compliance output stage.
Keywords :
active networks; bioelectric potentials; biomedical electronics; prosthetics; transistors; triodes; constant current stimulation; current driver; current-controlled stimulation; headroom voltage; high density implantable stimulators; implicit compliance monitor; output impedance; power efficient output stage; regulated cascode structure; transistor; triode region; voltage-controlled stimulation;
Journal_Title :
Electronics Letters
DOI :
10.1049/el.2012.0255