DocumentCode :
1515789
Title :
Automated Fiducial Localization in CT Images Based on Surface Processing and Geometrical Prior Knowledge for Radiotherapy Applications
Author :
Fattori, Giovanni ; Riboldi, Marco ; Desplanques, Maxime ; Tagaste, Barbara ; Pella, Andrea ; Orecchia, Roberto ; Baroni, Guido
Author_Institution :
Dept. of Bioeng., Politec. di Milano, Milano, Italy
Volume :
59
Issue :
8
fYear :
2012
Firstpage :
2191
Lastpage :
2199
Abstract :
We propose a novel method for radio-opaque external marker localization in CT scans for infrared (IR) patient set-up in radiotherapy. Efforts were focused on the quantification of uncertainties in marker localization in the CT dataset as a function of algorithm performance. We implemented a 3-D approach to fiducial localization based on surface extraction and marker recognition according to geometrical prior knowledge. The algorithm parameters were optimized on a clinical CT dataset coming from 35 cranial and extra-cranial patients; the localization accuracy was benchmarked at variable image resolution versus laser tracker measurements. The applicability of conventional IR optical tracking systems for localizing external surrogates in daily patient set-up procedures was also investigated in 121 proton therapy treatment sessions. Our study shows that the implemented algorithm features surrogates localization with uncertainties lower than 0.3 mm and with a true positive rate of 90.1%, being this latter mainly influenced by fiducial homogeneity in the CT images. The reported clinical validation in proton therapy confirmed the submillimetric accuracy and the expected algorithm sensitivity. Geometrical prior knowledge allows judging the reliability of the extracted fiducial coordinates, ensuring the highest accuracy in patient set-up.
Keywords :
computerised tomography; image recognition; medical image processing; radiation therapy; CT images; algorithm performance; automated fiducial localization; extracranial patients; geometrical prior knowledge; image resolution; laser tracker measurement; localization accuracy; marker recognition; optical tracking systems; radioopaque external marker localization; radiotherapy; surface extraction; surface processing; Accuracy; Computed tomography; Measurement by laser beam; Medical treatment; Surface morphology; Surface treatment; Three dimensional displays; Automated detection; CT scanning; fiducial localization; particle therapy; patient positioning; radiotherapy; Algorithms; Fiducial Markers; Humans; Imaging, Three-Dimensional; Phantoms, Imaging; Radiotherapy Planning, Computer-Assisted; Skull; Tomography, X-Ray Computed;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2012.2198822
Filename :
6198875
Link To Document :
بازگشت