DocumentCode :
1517368
Title :
Calibration of uniform-field spark-gaps for high-voltage measurement at power frequencies
Author :
Bruce, F.M.
Author_Institution :
British Electrical and Allied Industries Research Association, Leatherhead, UK
Volume :
94
Issue :
38
fYear :
1947
fDate :
4/1/1947 12:00:00 AM
Firstpage :
138
Lastpage :
149
Abstract :
A calibration is given for spark-gap electrodes designed to ensure that the spark path lies in a uniform field, and it is shown that for all three sizes of electrodes investigated, covering a range of 9 to 315 kV (peak), the sparkover voltages are within 0.2% of the values given by the empirical law V=24.22S+6.082¿S kV (peak) at 20°C and atmospheric pressure of 760 mm Hg, where S is the electrode spacing in centimetres. All observations were made with vertical gaps having the lower electrode connected to earth, and comparative data were obtained for a sphere gap with electrodes 25 cm in diameter. The effects of surrounding objects and of faulty alignment of the electrodes were determined experimentally, and on the basis of these results a number of recommendations are made, giving the limits within which the spark-over voltage is defined by the above equation.
Keywords :
spark gaps; voltmeters;
fLanguage :
English
Journal_Title :
Electrical Engineers - Part II: Power Engineering, Journal of the Institution of
Publisher :
iet
Type :
jour
DOI :
10.1049/ji-2.1947.0052
Filename :
5296671
Link To Document :
بازگشت