DocumentCode :
1518116
Title :
Optimization of Large-Angle Pinhole Collimator for Environmental Monitoring System
Author :
Baek, Cheol-Ha ; Lee, Seung-Jae ; Choi, Yong ; Chung, Yong Hyun
Author_Institution :
Dept. of Radiol. Sci., Yonsei Univ., Wonju, South Korea
Volume :
57
Issue :
3
fYear :
2010
fDate :
6/1/2010 12:00:00 AM
Firstpage :
1404
Lastpage :
1408
Abstract :
The purpose of this study was to optimize a large-angle pinhole collimator using Monte Carlo simulation for nuclear survey imaging. Simulations using GATE (Geant4 Application for Tomographic Emission) were performed to model the pinhole gamma camera system. A gamma camera consists of a cone-shaped pinhole collimator with a tungsten aperture and a CsI(Tl) scintillation crystal 6.0 mm thick and 50.0 mm × 50.0 mm in area. The focal length and the acceptance angle of the pinhole collimator were set to 14.5 mm and 120°, respectively. The intrinsic spatial resolution and sensitivity were simulated by changing the pinhole diameter and channel height. The point source of Tc-99m was located 30.0 mm above the center of the pinhole, and the projection data was estimated for pinhole diameter values from 0.5 mm to 4.0 mm while the channel heights were fixed between knife-edge and 3.0 mm. The optimal ranges of channel height and pinhole diameter were determined through evaluation of the intrinsic resolution and sensitivity tradeoff curves. The simulation results allowed us to determine the optimal values of pinhole diameter and channel height to be 1.5 mm and 0.5 mm, respectively, to get intrinsic resolution below 2.0 mm FWHM with a reasonable sensitivity for the system configured in this study. The resolution and sensitivity were measured experimentally, and the simulated and measured data were in good agreement. The results demonstrated that the pinhole collimator designed in this study could be utilized to create a large-angle radiation monitoring system.
Keywords :
Monte Carlo methods; collimators; environmental factors; gamma-ray detection; radiation monitoring; solid scintillation detectors; CsI(Tl) scintillation crystal; GATE; Geant4 Application for Tomographic Emission; Monte Carlo simulation; Tc-99m point source; channel height; cone shaped pinhole collimator; distance 30.0 mm; environmental monitoring system; intrinsic resolution curve; large angle pinhole collimator optimization; large angle radiation monitoring system; nuclear survey imaging; pinhole collimator acceptance angle; pinhole collimator focal length; pinhole diameter; pinhole gamma camera system; projection data; sensitivity tradeoff curve; size 0.5 mm; size 1.5 mm; size 50.0 mm; tungsten aperture; Apertures; Biomedical imaging; Collimators; Design optimization; Digital cameras; Monitoring; Optical imaging; Scattering; Spatial resolution; Tomography; Environmental monitoring system; Geant4 Application for Tomographic (GATE); gamma camera; large-angel pinhole collimator;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2009.2039877
Filename :
5485182
Link To Document :
بازگشت