DocumentCode
1518341
Title
Stability and characteristics of large CZT coplanar electrode detectors
Author
Pérez, José M. ; He, Zhong ; Webe, D.
Author_Institution
CIEMAT, Madrid, Spain
Volume
48
Issue
3
fYear
2001
fDate
6/1/2001 12:00:00 AM
Firstpage
272
Lastpage
277
Abstract
The single polarity charge sensing method based on coplanar grid electrodes has demonstrated promising results on CZT detectors having volumes of ~1 cm3. This paper presents the performance of two coplanar grid CZT detectors having dimensions of 1.5×1.5×1.0 cm2, among the largest CZT spectrometers that have been reported. Energy resolutions of 2.3% and 2.0 % full-width half-maximum were obtained on the two detectors at 662 keV using simple analog techniques. Further improvement is achievable with depth sensing and digital pulse processing techniques. Other detector characteristics, critical to practical application of these devices, are reported. This includes the stability of detector response over a period of several months and the variation of detector performance in environments having high gamma radiation fluxes and a range of counting rates
Keywords
electrodes; gamma-ray spectrometers; semiconductor counters; 662 keV; CZT spectrometers; CdZnTe; coplanar grid CZT detectors; depth sensing; digital pulse processing; energy resolution; full-width half-maximum; gamma ray; single polarity charge sensing method; Electrodes; Energy resolution; Gamma ray detection; Gamma ray detectors; Helium; Preamplifiers; Radiation detectors; Semiconductor radiation detectors; Spectroscopy; Stability;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.940064
Filename
940064
Link To Document