Electronic design of a multichannel programmable implant for neuromuscular electrical stimulation
Author
Arabi, Karim ; Sawan, Mohamad A.
Author_Institution
Dept. of Electr. Eng., Ecole de Technol. Superieure, Montreal, Que., Canada
Volume
7
Issue
2
fYear
1999
fDate
6/1/1999 12:00:00 AM
Firstpage
204
Lastpage
214
Abstract
An advanced stimulator for neuromuscular stimulation of spinal cord injured patients has been developed. The stimulator is externally controlled and powered by a single encoded radio frequency carrier and has four independently controlled bipolar stimulation channels. It offers a wide range of reprogrammability and flexibility, and can be used in many neuromuscular electrical stimulation applications. The implant system is adaptable to patient´s needs and to future developments in stimulation algorithms by reprogramming the stimulator. The stimulator is capable of generating a wide range of stimulation waveforms and stimulation patterns and therefore is very suitable for selective nerve stimulation techniques. The reliability of the implant has been increased by using a forward error detection and correction communication protocol and by designing the chip for structural testability based on scan test approach. Implemented testability scheme makes it possible to verify the complete functionality of the implant before and after implantation. The stimulators architecture is designed to be modular and therefore its different blocks can be reused as standard building blocks in the design and implementation of other neuromuscular prostheses. Design for low-power techniques have also been employed to reduce power consumption of the electronic circuitry
Keywords
biocontrol; biomedical electronics; neuromuscular stimulation; programmable controllers; prosthetics; advanced stimulator; bipolar stimulation channels; bladder controller; electronic design; externally controlled; flexibility; forward error correction; forward error detection; modular architecture; multichannel programmable implant; neuromuscular electrical stimulation; neuromuscular prostheses; reduced power consumption; reprogrammability; scan test approach; selective nerve stimulation; single encoded radio frequency carrier; spinal cord injured patients; structural testability; Electrical stimulation; Error correction; Implants; Neuromuscular stimulation; Prosthetics; Protocols; Radio control; Radio frequency; Spinal cord; Testing;
fLanguage
English
Journal_Title
Rehabilitation Engineering, IEEE Transactions on
Publisher
ieee
ISSN
1063-6528
Type
jour
DOI
10.1109/86.769411
Filename
769411
Link To Document