DocumentCode
1519193
Title
CCE 92-05 intercomparison of ac-dc voltage transfer standards at high frequencies (1-50 MHz)
Author
Van Mullem, Cees J.
Author_Institution
NMi Van Swinden Lab., Delft, Netherlands
Volume
48
Issue
2
fYear
1999
fDate
4/1/1999 12:00:00 AM
Firstpage
154
Lastpage
157
Abstract
From August 1995 to May 1998, the CCE 92-05 intercomparison of ac-dc voltage transfer standards at high frequencies was carried out. Two travel standards were measured by 15 national standards institutes. The results in the frequency range from 1 to 50 MHz show a good agreement between the majority of participants. The span of the majority of the reported ac-dc differences at 50 MHz is less than 1000 μV/V, which is similar to a previous intercomparison but with a greater number of participants
Keywords
transfer standards; voltage measurement; 1 to 50 MHz; CCE 92-05 intercomparison; high frequency AC-DC voltage transfer standard; thermal converter; travel standard; Calibration; Coaxial components; Connectors; Frequency; Helium; Instruments; Measurement standards; Particle measurements; Voltage; Wideband;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.769551
Filename
769551
Link To Document