    Results from a detailed calculation of the sensitivity of a cryogenic current comparator

  • Author

    Early, Murray D. ; Van Dam, Marcos A.

  • Author_Institution
    Meas. Stand. Lab., Ind. Res. Ltd., Lower Hutt, New Zealand
  • Volume
    48
  • Issue
    2
  • fYear
    1999
  • fDate
    4/1/1999 12:00:00 AM
  • Firstpage
    379
  • Lastpage
    382
  • Abstract
    A detailed calculation of the sensitivity of a cryogenic current comparator (CCC) has been completed. The model accounts for the current distributions on the closed external shield and the overlapped shield around the windings. Additional refinements to the model lead to excellent agreement (typically better than 5%) with sensitivity measurements for a wide range of CCC geometries. We demonstrate that generally the best strategy for significantly improving the overall gain of a CCC is to increase the number of windings in the toroid at the expense of a reduction in the sensitivity
  • Keywords
    SQUIDs; current comparators; current distribution; electric current measurement; sensitivity analysis; windings; SQUID connected; closed external shield; cryogenic current comparator; current distributions; inductance; model refinements; number of windings; overall gain improvement; overlapped shield; sensitivity calculation; small currents measurement; strip model; superconducting shielding; Cryogenic electronics; Current distribution; Electrical resistance measurement; Geometry; Hall effect; Inductance; Solid modeling; Strips; Superconducting coils; Superconducting filaments and wires;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.769605
  • Filename
    769605