DocumentCode
1519557
Title
Calorimetric thermal voltage converter as a wideband calculable standard of AC-DC difference
Author
Filipski, Piotr S. ; Clark, Richard F. ; Paulusse, David C.
Author_Institution
Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, Ont., Canada
Volume
48
Issue
2
fYear
1999
fDate
4/1/1999 12:00:00 AM
Firstpage
387
Lastpage
390
Abstract
A calorimetric thermal voltage converter is used at the National Research Council of Canada as a high-frequency reference standard of the Ac-DC transfer difference. This paper describes its redesigned realization with an optimized frequency characteristic. Results presented in the paper show that it can be used as a standard of AC-DC difference in the 1 Hz to 100 MHz frequency range. The agreement between the calculated and measured AC-DC transfer differences is better than ±10 μV/V at 10 MHz, and ±300 μV/V at 100 MHz
Keywords
calibration; calorimeters; convertors; thermopiles; transfer standards; voltage measurement; 1 Hz to 100 MHz; AC-DC difference; calibration; calorimetric thermal voltage converter; coaxial transmission lines; high-frequency reference standard; optimized frequency characteristic; thermopile output; transfer standard; uncertainty; wideband calculable standard; Analog-digital conversion; Coaxial components; Connectors; Councils; Frequency conversion; Resistors; Temperature; Testing; Voltage; Wideband;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.769607
Filename
769607
Link To Document