Title :
Automated comparator for accurate AC-DC difference measurements at the BNM-LCIE
Author :
Poletaeff, André
Author_Institution :
Lab. Central des Ind. Electr., Fontenay aux Roses, France
fDate :
4/1/1999 12:00:00 AM
Abstract :
This paper describes the new automated comparator system developed at the BNM-LCIE for the calibration of voltage thermal converters used as national AC-DC transfer standards. The system achieves Type A standard uncertainties of a few parts in 107 in the frequency range from 10 Hz to 1 MHz (100 kHz for voltages above 100 V). Type B standard uncertainties, depending on the type of converters to be compared and on the measurement voltage level, are between a few parts in 107 and a few parts in 106
Keywords :
calibration; comparators (circuits); convertors; measurement uncertainty; transfer standards; voltage dividers; voltage measurement; 10 Hz to 1 MHz; Type A standard uncertainties; accurate AC-DC difference measurement; automated comparator system; calibration; frequency response; nanovoltmeter; national AC-DC transfer standards; resistive dividers; type B standard uncertainties; voltage thermal converters; Analog-digital conversion; Automatic testing; Calibration; Frequency conversion; Frequency measurement; Frequency response; Impedance measurement; Measurement standards; Standards development; Voltage measurement;
Journal_Title :