• DocumentCode
    1519778
  • Title

    Accuracy evaluation of the optically pumped Cs frequency standard at NRLM

  • Author

    Hagimoto, Ken ; Ohshima, S. ; Nakadan, Y. ; Koga, Yasuki

  • Author_Institution
    Nat. Res. Lab. of Metrol., Ibaraki, Japan
  • Volume
    48
  • Issue
    2
  • fYear
    1999
  • fDate
    4/1/1999 12:00:00 AM
  • Firstpage
    496
  • Lastpage
    499
  • Abstract
    The optically pumped cesium (Cs) frequency standard at the National Research Laboratory of Metrology, Ibaraki, Japan (NRLM), NRLM-4, was modified in various ways in order to improve frequency stability and accuracy, We determined the frequency stability by comparing the phase with that of a hydrogen maser: the Allan deviation was σy(τ)=5×10-13/√τ. We estimated the frequency corrections and uncertainties due to various sources, and the total uncertainty was found to be 2.9×10-14 . The frequency difference between NRLM-4 and the international atomic time (TAI) was also measured
  • Keywords
    atomic clocks; frequency stability; frequency standards; measurement uncertainty; Allan deviation; Cs; NRLM-4; National Research Laboratory of Metrology; frequency corrections; frequency stability; international atomic time; optically pumped frequency standard; total uncertainty; Atomic measurements; Frequency estimation; Frequency measurement; Hydrogen; Laboratories; Masers; Metrology; Optical pumping; Stability; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.769640
  • Filename
    769640