DocumentCode
1519810
Title
Some results and an estimate of the current uncertainty of the PTBs reconstructed primary clock CS1
Author
Bauch, Andreas ; Fischer, Bernd ; Heindorff, Thomas ; Schröder, Roland
Author_Institution
Phys. Tech. Bundesanstalt, Braunschweig, Germany
Volume
48
Issue
2
fYear
1999
fDate
4/1/1999 12:00:00 AM
Firstpage
508
Lastpage
511
Abstract
The paper briefly describes the reconstructed primary clock CS1 of the Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany, the estimation of its type B standard uncertainty, and some results obtained during the first year of routine operation. The present estimate of the CS1 type B standard uncertainty is 7×10-15 (Iσ)
Keywords
atomic clocks; caesium; frequency standards; measurement uncertainty; Cs; PTB primary clock CS1; frequency standard; reconstructed primary clock; time standard; type B standard uncertainty; Atomic beams; Clocks; Frequency; Helium; Magnetic separation; Microwave ovens; Quantization; Resonance; Uncertainty; Vacuum systems;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.769644