Abstract
The national metrology laboratory, BNM-LCIE, France, has developed a set of four new waveguide primary thermal noise standards for the frequency band from 8.2 GHz to 40 GHz. Each noise standard is made with silicon carbide microwave absorber and inserted into a section of waveguide made in platinum 10% rhodium; the whole is put inside an electrical oven. The output noise temperature range of the standards obtained from temperature and attenuation measurements is within 948-1012 K±1.21.8 K, in one standard-type uncertainty. The calibration of gas discharge noise sources and/or solid-state noise sources is achieved by comparison to standards using a switching radiometer. The excess noise ratio given is within 15.35-15.65 dB for gas discharge sources and within 14-16 dB for solid-state sources, with typical uncertainty of 0.02 dB
Keywords
calibration; electric noise measurement; measurement standards; microwave measurement; rectangular waveguides; thermal noise; 8.2 to 40 GHz; Pt-Rh; SiC; attenuation measurement; calibration; electrical oven; electromagnetic metrology; gas discharge noise source; platinum-rhodium waveguide; primary thermal noise standard; silicon carbide microwave absorber; solid-state noise source; temperature measurement; Discharges; Frequency; Laboratories; Metrology; Microwave ovens; Platinum; Silicon carbide; Solid state circuits; Standards development; Temperature distribution;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.769676
Filename
769676
Link To Document