DocumentCode :
1520061
Title :
Developments in coaxial power standards at NPL
Author :
Ascroft, John T.
Author_Institution :
Centre for Electromagn. Metrol., Nat. Phys. Lab., Teddington, UK
Volume :
48
Issue :
2
fYear :
1999
fDate :
4/1/1999 12:00:00 AM
Firstpage :
647
Lastpage :
649
Abstract :
A 2.4 mm twin line coaxial calorimeter is the first microwave power standard in this line size to be developed at the National Physical Laboratory (NPL). The design is based on a twin line coaxial calorimeter where RF power dissipated in one load is measured in terms of a DC power dissipated in a similar load with its effective efficiency derived from a thermal and electrical analysis of the twin lines. The calorimeter will provide traceable power measurements from DC to 50 GHz between power levels of 0.1 and 10 mW with an uncertainty ranging from 0.5% at 10 MHz to 2.3% at 50 GHz at the 1 standard deviation level
Keywords :
calorimeters; coaxial cables; measurement standards; measurement uncertainty; microwave measurement; power measurement; power meters; 0.1 mW; 10 MHz to 50 GHz; 10 mW; 2.4 mm; DC power dissipated; NPL developments; RF power dissipated; coaxial power standards; effective efficiency; microwave power standard; terminated coaxial lines; traceable power measurements; twin line coaxial calorimeter; uncertainty; Bridge circuits; Coaxial components; Connectors; Laboratories; Power measurement; Radio frequency; Standards development; Temperature control; Temperature sensors; Thermal resistance;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.769678
Filename :
769678