Abstract
Time-varying magnetic fields can be used to transfer power across the skin to drive implantable biomedical devices without the use of percutaneous wires. However, the main challenges of a transcutanoues energy transfer (TET) system are the temperature rise caused by power loss in the implanted circuitry and the changes in positioning between the external and internal coils due to fitting and changes in posture. This study presents a TET system with a closed-loop frequency-based power regulation method to deliver the right amount of power to the load under variable coil coupling conditions. After implanting a TET system into adult sheep, the temperature rise in the internal and external coils of a TET system was measured for power delivery in the range of 5 W to 15 W. The sheep was housed in a temperature controlled (16 plusmn1degC, humidity 50plusmn10%) room, in accordance with the standard protocols implemented at the University of Auckland for sheep studies. A power-loss analysis for the overall system was performed. The system was capable of regulating power for axially aligned separations of up to 16 mm. The maximum power efficiency of the overall system was 82.1% and a maximum temperature rise of 2.7degC was observed on the implanted secondary coil.
Keywords
biomagnetism; biomedical equipment; coils; magnetic fields; prosthetic power supplies; closed-loop frequency-based power regulation; implantable biomedical device; power-loss analysis; time-varying magnetic field; transcutaneous energy transfer system; Biomedical measurements; Coils; Coupling circuits; Energy exchange; Frequency; Implantable biomedical devices; Magnetic fields; Skin; Temperature distribution; Wires; Magnetic filed; power efficiency; transcutaneous energy transfer (TET);
fLanguage
English
Journal_Title
Biomedical Circuits and Systems, IEEE Transactions on
Publisher
ieee
ISSN
1932-4545
Type
jour
DOI
10.1109/TBCAS.2009.2031539
Filename
5299164
Link To Document