Abstract :
A crucial issue of functional electrical stimulation (FES) is the control of motor function by the artificial activation of paralyzed muscles. Major problems that limit the success of current FES systems are the nonlinearity of the target system and the rapid change of muscle properties due to fatigue. In this study, four different strategies, including an adaptive algorithm, to control the movement of the freely swinging shank were developed on the basis of computer simulations and experimentally evaluated on two subjects with paraplegia due to a complete thoracic spinal cord injury. After developing a nonlinear, physiologically based model describing the dynamic behavior of the knee joint and muscles, an open-loop approach, a closed-loop approach, and a combination of both were tested. In order to automate the individual adjustments cited, we further evaluated the performance of an adaptive feedforward controller. The two parameters chosen for the adaptation were the threshold pulse width and the scaling factor for adjusting the active moment produced by the stimulated muscle to the fitness of the muscle. These parameters have been chosen because of their significant time variability. The first three controllers with fixed parameters yielded satisfactory result. An additional improvement was achieved by applying the adaptive algorithm that could cope with problems due to muscle fatigue, thus permitting on-line identification of critical parameters of the plant. Although the present study is limited to a simplified experimental setup, its applicability to more complex and functional movements can be expected.
Keywords :
adaptive control; biocontrol; biomechanics; closed loop systems; control nonlinearities; feedback; feedforward; identification; motion control; neuromuscular stimulation; nonlinear dynamical systems; patient rehabilitation; performance index; physiological models; piecewise linear techniques; three-term control; FES-induced single joint movements; PID controller; adaptive algorithm; adaptive feedforward controller; artificial activation; biomechanical model; closed-loop approach; computer simulations; dynamic behavior; feedback; freely swinging shank; knee joint; model-based control; motor function control; muscle fatigue; nonlinear physiologically based model; on-line identification; open-loop approach; paralyzed muscles; paraplegia; performance index; piecewise linear recruitment function; scaling factor; target system nonlinearity; thoracic spinal cord injury; threshold pulse width; Adaptive algorithm; Computer simulation; Fatigue; Knee; Muscles; Neuromuscular stimulation; Open loop systems; Programmable control; Spinal cord injury; Testing; Biomechanics; Computer Simulation; Electric Stimulation Therapy; Humans; Isometric Contraction; Knee Joint; Motor Neurons; Muscle Fatigue; Muscle, Skeletal; Paraplegia; Prosthesis Design; Range of Motion, Articular; Spinal Cord Injuries;
fLanguage :
English
Journal_Title :
Neural Systems and Rehabilitation Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
1534-4320
Type :
jour
DOI :
10.1109/7333.948452
Filename :
948452
Link To Document :
بازگشت