DocumentCode :
1527502
Title :
A Microfabricated Phantom for Quantitative MR Perfusion Measurements: Validation of Singular Value Decomposition Deconvolution Method
Author :
Ebrahimi, Behzad ; Swanson, Scott D. ; Chupp, Timothy E.
Author_Institution :
Dept. of Biomed. Eng., Univ. of Michigan, Ann Arbor, MI, USA
Volume :
57
Issue :
11
fYear :
2010
Firstpage :
2730
Lastpage :
2736
Abstract :
A perfusion phantom with unique features and a wide variety of applications in magnetic resonance imaging (MRI) and other imaging modalities is presented. Using microfabrication technique, a network of microchannels, in the scale of actual microvasculature, was created. The geometry of the network was determined based on Murray´s “minimum work” law to simulate the hemodynamic in actual capillary networks. The perfusion-related parameters, such as flow, volume ratio, and the transit time, were precisely calculated using a finite-element method based program. These parameters were also estimated through the deconvolution of the residue function from the tissue concentration-time curve in the perfusion model. The widely accepted singular value decomposition (SVD) method in standard sSVD and reformulated rSVD forms were used for the purpose of the deconvolution and regularization. The accuracy of these methods in the presence of delay and dispersion was investigated. Comparing the estimated values to the true values, the contribution of each of these sources of error to the total error in the estimated perfusion parameters was determined.
Keywords :
biomedical MRI; blood flow measurement; deconvolution; finite element analysis; haemorheology; phantoms; singular value decomposition; capillary network; finite-element method; hemodynamic; microfabricated phantom; microfabrication; microvasculature; perfusion phantom; quantitative MR perfusion measurement; residue function; singular value decomposition; singular value decomposition deconvolution method; tissue concentration-time curve; Biomedical imaging; blood flow measurement; inverse problems; magnetic resonance imaging (MRI);
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2010.2055866
Filename :
5499031
Link To Document :
بازگشت