DocumentCode :
1528590
Title :
Verification of the westervelt equation for focused transducers
Author :
Jing, Yun ; Shen, Du ; Clement, Greg T.
Author_Institution :
Dept. of Radiol., Brigham and Women´´s Hosp., Boston, MA, USA
Volume :
58
Issue :
5
fYear :
2011
fDate :
5/1/2011 12:00:00 AM
Firstpage :
1097
Lastpage :
1101
Abstract :
This study investigates the validity of the Westervelt equation for focused transducers. The angular spectrum method is employed to analyze the second-harmonic acoustic field under the weakly nonlinear approximation. Although it is well known that the Westervelt equation is accurate for the case of quasi-plane waves, the present work demonstrates accurate solution for the highly focused case of a spherically-curved ultrasound transducer having an aperture angle of 80°. It is further found that the solution error is inversely dependent on the nonlinearity coefficient.
Keywords :
harmonic generation; ultrasonic transducers; Westervelt equation; angular spectrum method; focused transducer; quasi plane wave; second harmonic acoustic field; spherically curved ultrasound transducer; weakly nonlinear approximation; Acoustics; Apertures; Equations; Harmonic analysis; Mathematical model; System-on-a-chip; Transducers;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2011.1910