DocumentCode
1531253
Title
Low-leakage wide-access magnet for MRI
Author
Momy, A. ; Taquin, J.
Author_Institution
Inst. d´´Electron. Fondamentale, CNRS, Orsay, France
Volume
33
Issue
6
fYear
1997
fDate
11/1/1997 12:00:00 AM
Firstpage
4572
Lastpage
4574
Abstract
We propose a novel magnet structure for magnetic resonance imaging. The design made use of the CALMAG3D code which features high precision and high speed calculation and requires relatively small computer memory. The magnetic field at any point in space can be calculated without using the classical methods of finite elements, conformal transform, boundary limit conditions, and matrix inversion
Keywords
biomedical NMR; electrical engineering computing; electromagnets; magnetic fields; permanent magnets; variational techniques; CALMAG3D code; MRI; coils; ferromagnetic yokes; high precision high speed calculation; hybrid magnet; low-leakage wide-access magnet; magnet structure; magnetic field calculation; magnetic resonance imaging; permanent magnets; software simulation code; variational method; whole body medical imaging; Coils; Costs; Energy consumption; Finite element methods; Magnetic fields; Magnetic resonance imaging; Magnetic shielding; Magnetostatics; Permanent magnets; Superconducting magnets;
fLanguage
English
Journal_Title
Magnetics, IEEE Transactions on
Publisher
ieee
ISSN
0018-9464
Type
jour
DOI
10.1109/20.649897
Filename
649897