DocumentCode :
1531259
Title :
Development of a high speed X-ray tomography system for multiphase flow imaging
Author :
Morton, E.J. ; Luggar, R.D. ; Key, M.J. ; Kundu, A. ; Távora, L. M N ; Gilboy, W.B.
Author_Institution :
Dept. of Phys., Surrey Univ., Guildford, UK
Volume :
46
Issue :
3
fYear :
1999
fDate :
6/1/1999 12:00:00 AM
Firstpage :
380
Lastpage :
384
Abstract :
This paper describes a new system for quantitative measurement of multiphase flow. It is based on the use of high speed (50 frames per second) X-ray tomographic imaging. This requires a new X-ray tube concept to be implemented, together with an annular detector array and data acquisition system. A pressurised krypton microgap detector is being developed to give 90% quantum efficiency for detection of Kα radiation from the silver anode X-ray tube. Results from simulation work are presented to justify the overall design strategy selected
Keywords :
X-ray imaging; data acquisition; flow visualisation; multiphase flow; tomography; X-ray tomographic imaging; annular detector array; data acquisition system; high speed X-ray tomography system; multiphase flow imaging; pressurised krypton microgap detector; silver anode X-ray tube; Data acquisition; Fluid flow measurement; Optical imaging; Radiation detectors; Sensor arrays; Silver; X-ray detection; X-ray detectors; X-ray imaging; X-ray tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.775548
Filename :
775548
Link To Document :
بازگشت